Measuring principle: |
UV-Absorption |
Wavelength: |
253.7 nm |
UV source: |
electrodeless low-pressure mercury lamp |
Stabilization method: |
reference beam feedback lamp control method and thermal stabilisation |
Optical cell: |
fused silica (Suprasil), 23 cm long |
Optical cell heating: |
approx. 50°C |
Measuring range: |
lowest range: 0.1...100 µg/m³ (0.01-10 ppb)
highest range: 0...2000 ug/m³ (0-200 ppb) |
Sensitivity: |
0.1 µg/m³ (0,01 ppb) |
Response time (99% signal): |
< 1 second |
Computation of mean value: |
automatically via three freely selectable times |
Alarm: |
when concentration is exceeded, 3 levels programmable |
Status alarms: |
measuring cell soiled, battery state, UV source |
Control pad: |
waterproof membrane keypad |
Measurement Display: |
graphical LC display with background illumination |
Signal outputs: |
4...20 mA for recorder;
RS 232 bidirectional for PC, data logger etc.;
parallel for printer |
Pump: |
membrane pump, approx. 2 L / min |
Filter: |
PTFE, 1 µ, 47-50 mm diameter |
Power Supply: |
230 V / 50 Hz (115 V / 60 Hz option) |
Battery operation: |
external 12 DC source; optional 12 V batteries integrated, ca. 5h capacity |
Power consumption: |
30 VA |
Dimensions: |
45 x 15 x 35 cm (WxHxD) |
Weight: |
approx. 7 kg |